Back to the results guide

Bioimpedance: from signal to estimate

InBody does not directly measure kilograms of fat or muscle. It records weight and the electrical response of the body; water, fat-free mass, muscle and fat emerge from the subsequent model.

Where to find it: The impedance table by frequency and segment among the additional parameters.

Illustration of multi-frequency current paths through body segments

01

1. The device measures electrical opposition across segments and frequencies

A weak alternating current passes between contact electrodes, and the device records resistance and reactance separately for the arms, trunk and legs.

Resistance mainly relates to the conductive path through body fluids. Reactance captures the time shift at cell membranes. Together they form impedance; their relationship is linked to phase angle.

Multiple frequencies and segmental measurement add information about how the electrical response differs between body regions and between low and high frequencies. Impedance itself, however, is not yet a litre of water or a kilogram of muscle.

  • Measured: weight, resistance and reactance.
  • Calculated from the signal: phase angle and modelled limiting values.
  • An electrical number is neither good nor bad by itself without context.
Electrical signal
Impedance and reactance
Prediction model
Body-composition estimate
Teaching diagram — these are not your measured data.

02

2. Body composition emerges from the subsequent model

From the electrical response, the device first derives water distribution and only then other body components.

The simplified chain is impedance → body water → fat-free mass → fat as the difference from body weight. Skeletal muscle mass and many other items are further model outputs, not separate direct measurements.

The manufacturer does not publish every equation, so the result cannot be independently recalculated from the raw values. An estimate can be useful and repeatable, but the precision of one number is not the same as the accuracy of a laboratory reference method.

  • Body water is the first interpretive layer above the electrical signal.
  • Fat-free mass is not a synonym for skeletal muscle mass.
  • Fat mass emerges in the model as a difference, not by directly scanning fat.

03

3. Measurement conditions affect the whole chain

When fluid distribution or electrode contact changes, the input signal shifts and derived values can shift with it.

For a trend, keep a similar time of day, interval from food, drink, training and sauna, and correct hand and foot contact with the electrodes. A short-term change need not represent a true tissue change.

A series of measurements on the same device under comparable conditions provides the most value. When comparing with DEXA, another BIA scale or anthropometry, remember that the methods use different physical inputs and models.

Evidence used

The guide follows manufacturer documentation, consensus BIA principles and expert reviews of the method. It consistently separates the raw electrical signal, subsequent derivation and the resulting body-composition model.

Content reviewed · 2026-07-26